Re: problem with harvesting cells after sorting
(cawort01@roentgen.bcc.louisville.edu)
Fri, 15 Mar 96 10:55:15 EST
what are you sorting the cells into? if you are using plastic tubes you are getting what I jokingly call "like-Liked". the tube will gradually pick up the charge of the droplets. and when that charge gets high enough your sorted cells will be repelled by the tube.(not a good situation<G>). all you need to do is use glass tubes. I hope this helps.
Chris
Connecting@Warp.SPEED