RE: flow assay for hereditary spherocyto

Leary, James (jleary@mspo3.med.utmb.edu)
Wed, 26 Mar 1997 08:26:00 -0600

I would not expect light scatter pulse width to work. However, a
combination of forward scatter, side scatter and pulse rise-time
time-of-flight should work. Pulse rise-time is sensitive to the shape,
not just the diameter, of cells. If you have access to this measurement
on your flow cytometer, use start and stop thresholds of 5% and 35%
respectively for greatest sensitivity.

-----Original Message-----
From: Yashpal Agrawal [SMTP:yagrawal@messi.uku.fi]
Sent: Tuesday, March 25, 1997 5:43 PM
To: cyto-inbox
Subject: flow assay for hereditary spherocytosis

I tried to measure pulse width on a FACStar machine for this purpose. The
method was not able to differentiate from normal RBC's.
Hope this is useful.
Yash Pal Agrawal
Kuopio, Finland


Home Page Table of Contents Sponsors E-Mail Archive Web Sites

CD-ROM Vol 3 was produced by Monica M. Shively and other staff at the Purdue University Cytometry Laboratories and distributed free of charge as an educational service to the cytometry community. If you have any comments please direct them to Dr. J. Paul Robinson, Professor & Director, PUCL, Purdue University, West Lafayette, IN 47907. Phone: (765)-494-0757; FAX(765) 494-0517; Web http://www.cyto.purdue.edu , EMAIL cdrom3@flowcyt.cyto.purdue.edu