Re: problem with harvesting cells after sorting
(cawort01@roentgen.bcc.louisville.edu)
Fri, 15 Mar 96 10:55:15 EST
what are you sorting the cells into? if you are using plastic tubes you are getting what I jokingly call "like-Liked". the tube will gradually pick up the charge of the droplets. and when that charge gets high enough your sorted cells will be repelled by the tube.(not a good situation<G>). all you need to do is use glass tubes. I hope this helps.
Chris
Connecting@Warp.SPEED
CD-ROM Vol 3 was produced by Monica M. Shively and other staff at the
Purdue University Cytometry Laboratories and distributed free of charge
as an educational service to the cytometry community.
If you have any comments please direct them to
Dr. J. Paul Robinson, Professor & Director,
PUCL, Purdue University, West Lafayette, IN 47907.
Phone: (765)-494-0757;
FAX(765) 494-0517;
Web
http://www.cyto.purdue.edu
, EMAIL
cdrom3@flowcyt.cyto.purdue.edu