leaky FACScan supply tanks

vanburen%flovax.dnet@rocdec.roc.wayne.edu
Wed, 19 May 93 18:41:03 -0400

MAK (alias Mark KuKuruga) and others:

We've had leaky tank problems, too. The most recent was when the back seam
broke. Not a big hole, mind you, but a nasty slow drip! After checking all
sorts of seals, tubing, filters and such first (and wasting lots of time), I
found the leak in the tank. Response from BD: consumable item, send $7.02
for a replacement, and have a nice day!

As for cap leaks, I've had to modify the supply tank in order to (hopefully)
stop the leaks.

Eric Van Buren vanburen%flovax.dnet@rocdec.roc.wayne.edu
Immunology & Microbiology 313 577 1009 voice
Wayne State University 313 577 1155 fax


Home Page Table of Contents Sponsors E-Mail Archive Web Sites

CD-ROM Vol 3 was produced by Monica M. Shively and other staff at the Purdue University Cytometry Laboratories and distributed free of charge as an educational service to the cytometry community. If you have any comments please direct them to Dr. J. Paul Robinson, Professor & Director, PUCL, Purdue University, West Lafayette, IN 47907. Phone: (765)-494-0757; FAX(765) 494-0517; Web http://www.cyto.purdue.edu , EMAIL cdrom3@flowcyt.cyto.purdue.edu