Facscan side scatter problem

Vera Svobodova (Marcia.Woda@BANYAN.UMMED.EDU)
Thu, 21 Dec 95 10:19:57 EST

I agree with M Kukuruga. This happens very rarely to us and usually in whole
blood specimens. I have always felt that it was a wad of something going
through overwhelming the threshold trigger. It is annoying and bad when it
happens during a crucial specimen, but we abort and restart acquisition
rather than trying to gate it out as percentages will be affected.


Home Page Table of Contents Sponsors Web Sites
CD ROM Vol 2 was produced by staff at the Purdue University Cytometry Laboratories and distributed free of charge as an educational service to the cytometry community. If you have any comments please direct them to Dr. J. Paul Robinson, Professor & Director, PUCL, Purdue University, West Lafayette, IN 47907. Phone:(317) 494-0757; FAX (317) 494-0517; Web http://www.cyto.purdue.edu EMAIL robinson@flowcyt.cyto.purdue.edu