bleach anecdote
Howard Shapiro (Marcia.Woda@BANYAN.UMMED.EDU)
Thu, 13 Feb 97 17:36:46 EST
To add to the clean-machine thread, monthly exposure of the fluidics on a
Calibur to a warm 10% bleach solution over a year's time undoubtedly led to
the demise of a 15$ plastic part that resulted in an 850$ service call. The
flow cell however is sparkly clean ...
CD-ROM Vol 3 was produced by Monica M. Shively and other staff at the
Purdue University Cytometry Laboratories and distributed free of charge